Equilibration time depends on the metal, adsorbate, and temperature. For some systems equilibration is complete within minutes. For activated chemisorption systems such as H2 on cobalt at ambient temperature, equilibration may require hours, or the measurement must be performed at elevated temperature to achieve reasonable kinetics. If the isotherm is collected before true equilibrium is reached at each pressure point, the measured uptake will be lower than the true equilibrium value. This causes the calculated crystallite size to be larger than the actual value and the calculated dispersion to be lower. For a new metal-adsorbate system, the only reliable way to confirm equilibration is to vary the equilibration time per point and verify that the measured uptake does not increase with additional waiting time. See our chemisorption analysis adsorbate selection guide for guidance on equilibration conditions for common catalyst systems.